MAUDE Adverse Event Report: ALLERGAN LAP-BAND AP ADJUSTABLE GASTRIC BANDING SYSTEM (STANDARD)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN LAP-BAND AP ADJUSTABLE GASTRIC BANDING SYSTEM (STANDARD) Back to Search Results
Catalog Number B-2240
Device Problem Detachment of Device or Device Component (2907)
Patient Problem Failure of Implant (1924)
Event Date 01/15/2015
Event Type  malfunction  
Event Description
Pt reported they have "never been able to reach satiety [with their lap-band (r) system], they've maxed out the fluid in the band that they could put in it." the doctor feels the lap-band(r) is "unbuckled or broken in the back" and it's "in a c shape with all the fluid in it, and then drops back when it's not." the lap- band(r) system remains implanted at this time.
 
Manufacturer Narrative
Taper ii.The reporter of the complaint was asked to return the product for analysis if it is explanted in the future.The device was repositioned and remains implanted.Based upon the model number, serial number and implant date provided by the reporter, the connector type is assumed to be a taper ii.Visual examination may determine the connector type associated with this report.The reporter of the event was asked to return the product for analysis, if it is explanted in the future.The surgery has not occurred, so allergan has not received the device nor performed analysis at this time.Device labeling addresses the possible outcome of an unbuckled band as follows: "the mfr of the lap-band ap (tm) adjustable gastric banding system has designed, tested and manufactured it to be reasonably fit for its intended use.However, the lap-band ap (tm) system is not a lifetime product and it may break or fail, in whole or in part, at any time after implantation and notwithstanding the absence of any defect.Causes of partial or complete failure include, without limitation, expected or unexpected bodily reactions to the presence and position of the implanted device, rare or atypical med complications, component failure and normal wear and tear.In addition, the lap-band ap(tm) system may be easily damaged by improper handling or use.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAP-BAND AP ADJUSTABLE GASTRIC BANDING SYSTEM (STANDARD)
Manufacturer (Section D)
ALLERGAN
la aurora de heredia
CS 
Manufacturer (Section G)
COSTA RICA
900 parkway global park zona franca
la aurora de heredia
CS  
Manufacturer Contact
krista alvarado
301 w. howard ln
ste 100
austin, TX 78753
5128132638
MDR Report Key4502623
MDR Text Key5348793
Report Number2024601-2015-00022
Device Sequence Number1
Product Code LTI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P000008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Company Representative
Reporter Occupation Not Applicable
Type of Report Initial
Report Date 01/15/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/03/2015
Device Catalogue NumberB-2240
Device Lot Number2455243
Other Device ID Number(01)10811955020183
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/15/2015
Initial Date FDA Received02/06/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/01/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age53 YR
-
-