GE OEC MEDICAL SYSTEMS (SLC) 9800; FLUOROSCOPIC X-RAY Back to Search Results
Model Number 9800
Device Problems No Display/Image (1183); Loss of Data (2903)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/19/2014
Event Type  malfunction  
Event Description
It was reported that the system had an intermittent loss of live video.This resulted in an intermittent, recoverable loss of imaging functionality.There is no report od injury or death associated with the event described in this complaint.
 
Manufacturer Narrative
A ge service rep performed an on site investigation.The failure could not be duplicated.As a precaution, the image intensifier power supply was replaced.The system was tested and found to be working as intended and put back into service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
9800
Type of Device
FLUOROSCOPIC X-RAY
Manufacturer (Section D)
GE OEC MEDICAL SYSTEMS (SLC)
384 wright brothers dr.
salt lake city UT 84116
Manufacturer (Section G)
GE OEC MEDICAL SYSTEMS
384 wright brothers dr.
salt lake city UT 84116
Manufacturer Contact
384 wright brothers dr.
salt lake city, UT 84116
8015364952
MDR Report Key4515720
MDR Text Key5532254
Report Number1720753-2014-08476
Device Sequence Number1
Product Code JAA
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 10/08/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9800
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/19/2014
Initial Date FDA Received10/08/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-