Device Problem Inaccurate Delivery (2339)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/14/2015
Event Type  Injury  
Event Description
The customer reported that while in patient use the respiratory therapist noticed the ventilator driver had an irregular beat and that it sort of hesitated and then resumed the fast pace.No patient compromise; the patient remained stable.The patient was removed from the ventilator and placed on another 3100b, no patient harm.
 
Manufacturer Narrative
(b)(4).The customer evaluated the ventilator and verified the complaint.However, he noticed that after an hour or more, the driver didn't "hesitate" or change a beat.When he performed the performance check, the amplitude was really low.The customer replaced the driver power module to fix the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAREFUSION
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
CAREFUSION
yorba linda CA
Manufacturer (Section G)
CAREFUSION
1100 bird center dr.
palm springs CA 92262
Manufacturer Contact
wendy schumacher
7607787219
MDR Report Key4532482
MDR Text Key16556856
Report Number2021710-2015-00288
Device Sequence Number1
Product Code LSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 01/14/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100B
Device Catalogue Number770155
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/14/2015
Initial Date FDA Received02/13/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/01/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-