The following description of the event was copied by a carefusion technical support specialist from an email from the distributor in (b)(6)."obf blender failed pre-delivery inspection.Blender is reading high on 21% o2 values (29%), and low on 600% o2 values (54.7%).Unable to resolve through calibration.Values for air and oxygen inlet pressures at 50psi: 21% = 29, 30% = 29.2, 60% = 54.7, 90% = 90.4, 100% = 100".
|
Failure analysis: microblender pn: (b)(4), sn: (b)(4) was received into the carefusion failure analysis lab for investigation.The carefusion failure analysis lab technician examined the microblender and found that it is out of calibration.The microblender was tested on test station # (b)(4) per (b)(4) and found that it is out of calibration at 21%, 30%, and 60% o2.At end point calibration sequence # 1, 21%, spec.Is 20.9 to 22.9% reads 25.8%.At set point calibration sequence # 2, 30%, spec.Is 28.0 to 32.0% reads 25.8%.At set point calibration sequence # 3, 60%, spec.Is 58.0 to 62.0% reads 53.4%.At set point calibration sequence # 7, 60%, spec.Is 58.0 to 62.0% reads 53.4%.At set point calibration sequence # 8, 60%, spec.Is 57.0 to 63.0% reads 53.5%.At set point calibration sequence # 9, 60%, spec.Is 57.0 to 63.0% reads 53.2%.Additional issues related to complaint; failed, at ¿flow rate¿ sequence # 12, spec.Is >/=30.0 lpm min, reads 23.Failed, at ¿flow rate¿ sequence # 15, spec.Is >/=30.0 lpm min, reads 23.The microblender was recalibrated by adjusting the valve seat and the control knob.Finding/root-cause: duplicate, the microblender is out of calibration, complaint allegation.
|