• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MIDMARK CORP JB-70; UNIT, X-RAY, EXTRAORAL WITH TIMER, PRODUCT CODE: EHD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MIDMARK CORP JB-70; UNIT, X-RAY, EXTRAORAL WITH TIMER, PRODUCT CODE: EHD Back to Search Results
Model Number JB-70
Device Problem Installation-Related Problem (2965)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/08/2008
Event Type  malfunction  
Event Description
It was reported to midmark that an x-ray unit (jb-70 intraoral dental x-ray system, serial (b)(4)) separated from its wall mount in a user facility.No injuries were reported.Root cause was determined to be installation error due to the mounting bolt not properly engaging with the wall stud.The bolt had been inserted at an angle that made it exit out the side of the wall stud.This mdr is being submitted as part of retrospective complaint at midmark corporation ((b)(6) midmark (formerly progeny) began filing mdrs for x-ray units separating from their wall mounts in (b)(6) 2009 (refer to mdr 1423380-2009-00006).All subsequent incidents involving units fully separating from the wall mount have been submitted to cdrh as mdrs.However, as part of retrospective complaint review, midmark has identified this unreported incident that occurred prior to (b)(6) 2009 and has decided to report this incident for consistency.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JB-70
Type of Device
UNIT, X-RAY, EXTRAORAL WITH TIMER, PRODUCT CODE: EHD
Manufacturer (Section D)
MIDMARK CORP
675 heathrow drive
lincolnshire IL 60069
Manufacturer Contact
nataliya khutorna
675 heathrow drive
lincolnshire, IL 60069
8474159760
MDR Report Key4546531
MDR Text Key5517628
Report Number1423380-2015-00008
Device Sequence Number1
Product Code EHD
Combination Product (y/n)N
PMA/PMN Number
K020070
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Distributor
Reporter Occupation Other
Type of Report Initial
Report Date 09/08/2008
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/23/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberJB-70
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received09/08/2008
Was Device Evaluated by Manufacturer? No
Date Device Manufactured02/01/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-