PENUMBRA, INC. PENUMBRA SYSTEM 5MAX ACE REPERFUSION CATHETER; NRY Back to Search Results
Catalog Number 5MAXACE132
Device Problems Aspiration Issue (2883); Infusion or Flow Problem (2964)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/05/2014
Event Type  malfunction  
Event Description
The patient was undergoing a thrombectomy procedure in the middle cerebral artery using a penumbra system 5max ace reperfusion catheter.During the procedure, the physician claimed that flow was limited through the ace catheter and was unable to aspirate.The ace catheter was repositioned and the physician was still unable to aspirate the thrombus.The ace catheter was removed from the patient and the procedure successfully continued using a new penumbra system 5max ace reperfusion catheter.There was no report of an adverse effect on the patient.
 
Manufacturer Narrative
Conclusion: the product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.The hospital discarded the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SYSTEM 5MAX ACE REPERFUSION CATHETER
Type of Device
NRY
Manufacturer (Section D)
PENUMBRA, INC.
1351 harbor bay parkway
alameda CA 94502
Manufacturer (Section G)
PENUMBRA, INC.
1351 harbor bay parkway
alameda CA 94502
Manufacturer Contact
kathleen kidd
1351 harbor bay parkway
alameda, CA 94502
5107483200
MDR Report Key4552469
MDR Text Key22150838
Report Number3005168196-2015-00185
Device Sequence Number1
Product Code NRY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K090752
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 01/27/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date08/31/2017
Device Catalogue Number5MAXACE132
Device Lot NumberF45703
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/27/2015
Initial Date FDA Received02/26/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/29/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age28 YR
-
-