• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES POWER TOOLS MINIMAL ACCESS ATTACHMENT; DRILLS, BURRS, TREPHINES AND ACCESSORIES (SIMPLE, POWERED)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES POWER TOOLS MINIMAL ACCESS ATTACHMENT; DRILLS, BURRS, TREPHINES AND ACCESSORIES (SIMPLE, POWERED) Back to Search Results
Catalog Number MA-D20
Device Problem Difficult To Position (1467)
Patient Problem No Patient Involvement (2645)
Event Date 02/06/2015
Event Type  malfunction  
Event Description
It was reported that during service and repair pre-testing, it was observed that the attachment device was heating in excess.The event was not related to surgery.There was no patient involvement reported.There were no injuries or medical intervention associated with this event.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
The actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and the reported condition was confirmed.An assessment was performed on the device which found that there was excess heat due to excessive corrosion on the internal components which was indicative of not following emersion / sterilization procedures.The assignable root cause was determined to be due to component damage caused by misuse, abuse and/ or user error.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINIMAL ACCESS ATTACHMENT
Type of Device
DRILLS, BURRS, TREPHINES AND ACCESSORIES (SIMPLE, POWERED)
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
rodney crain
4500 riverside drive
palm beach gardens, FL 33410
5616271080
MDR Report Key4564973
MDR Text Key5554869
Report Number1045834-2015-10482
Device Sequence Number1
Product Code HBE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK042783
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Service and Testing Personnel
Type of Report Initial,Followup
Report Date 02/06/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/03/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberMA-D20
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/05/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received03/10/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/27/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-