The cause for the advia centaur xp ca 125ii observed low qc level 2 recovery and patient results may have been attributed to specific reagent readypack related issue.The new reagent readypack was from the same reagent lot and from the same original reagent shipment to the customer.Siemens checked advia centaur readypacks reagent lot 015163 retains and observed normal recovery.The original readypack had been in use by the customer for a few hours, therefore an onboard reagent stability issue is unlikely.The original reagent readypack will not be returned to siemens for further investigation.No conclusion can be drawn.The instructions for use (ifu) states the following under the limitations section: "note: do not interpret levels of ca 125 as absolute evidence of the presence or the absence of malignant disease.Before treatment, patients with confirmed ovarian carcinoma frequently have levels of ca 125 within the range observed in healthy individuals.Elevated levels of ca 125 can be observed in patients with nonmalignant diseases.Measurements of ca 125 should always be used in conjunction with other diagnostic procedures, including information from the patient's clinical evaluation.The concentration of ca 125 in a given specimen determined with assays from different manufacturers can vary due to differences in assay methods, calibration, and reagent specificity.Ca 125 determined with different manufacturers' assays will vary depending on the method of standardization and antibody specificity." the instrument is performing with specifications.No further evaluation of the device is required.
|