Model Number 3716 |
Device Problems
Device Stops Intermittently (1599); Low impedance (2285); Impedance Problem (2950)
|
Patient Problem
Inadequate Pain Relief (2388)
|
Event Date 02/10/2015 |
Event Type
Injury
|
Event Description
|
Device 1 of 2.Reference mfr.Report# 1627487-2015-20145.
|
|
Event Description
|
Device 1 of 2.Reference mfr.Report# 1627487-2015-20145.
|
|
Manufacturer Narrative
|
Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
|
|
Manufacturer Narrative
|
Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
|
|
Event Description
|
Device 1 of 2.Reference mfr.Report# 1627487-2015-20145.It was reported the patient experienced intermittent stimulation.Diagnostics determined invalid /low impedances on the scs system.Consequently, the scs system was explanted.The patient received two leads with the same lot number.
|
|
Manufacturer Narrative
|
(b)(4).Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
|
|