DEPUY SYNTHES POWER TOOLS BATTERY REAMER/DRILL FOR BPL II; MOTOR, SURGICAL INSTRUMENT, AC-POWERED
|
Back to Search Results |
|
Catalog Number 530.705 |
Device Problem
Failure to Shut Off (2939)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 02/08/2015 |
Event Type
malfunction
|
Event Description
|
It was reported that during pre-surgery, it was observed that the battery reamer/drill device continued to run after the trigger had been released.There were no delays in the surgical procedure as an identical spare device was available for use.There was no patient involvement reported.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Manufacturer Narrative
|
The actual device was returned for evaluation.Reliability engineering evaluated the device and the reported condition was not duplicated nor confirmed.Therefore, an assignable root cause was not determined.However, during evaluation, it was observed that the device failed torque test.The assignable root cause was determined to be due to various worn components and seal deterioration from immersion during cleaning.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|