Catalog Number 530.705
Device Problem Failure to Shut Off (2939)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/08/2015
Event Type  malfunction  
Event Description
It was reported that during pre-surgery, it was observed that the battery reamer/drill device continued to run after the trigger had been released.There were no delays in the surgical procedure as an identical spare device was available for use.There was no patient involvement reported.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
The actual device was returned for evaluation.Reliability engineering evaluated the device and the reported condition was not duplicated nor confirmed.Therefore, an assignable root cause was not determined.However, during evaluation, it was observed that the device failed torque test.The assignable root cause was determined to be due to various worn components and seal deterioration from immersion during cleaning.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BATTERY REAMER/DRILL FOR BPL II
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH44 37
SZ  CH4437
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH44 37
SZ   CH4437
Manufacturer Contact
linda plews
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key4573704
MDR Text Key5555108
Report Number3009450871-2015-10534
Device Sequence Number1
Product Code GEY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 02/09/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/05/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number530.705
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/25/2015
Date Manufacturer Received02/09/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/28/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-