Model Number 3711
Device Problem Overheating of Device (1437)
Patient Problem No Information (3190)
Event Date 02/09/2015
Event Type  Injury  
Event Description
Device 1 of 2.Reference mfr.Report#: 1627487-2015-05107.It was reported the patient (b)(6) experienced heating while recharging their ipg.A replacement charging system will be sent to the patient to address the issue.On 08/01/2012, st.Jude medical, neuromodulation division, sent field action letters to patients related to heating while charging and raised awareness of this issue to patients.An increase in prior non-reported heating while charging events and other non-reported events was expected.
 
Event Description
Device 1 of 2.Reference mfr.Report#: 1627487-2015-05107.
 
Manufacturer Narrative
This charger model was associated with a field correction.Manufacturer's evaluation: corrective and preventive action capa) investigation was performed.Result - pocket heating confirmed.The investigation for capa (b)(4) associated with heating while charging (pocket heating) concluded that the charger was capable of transferring energy to the ipg at a rate that would cause heating of the ipg and/or charging wand of sufficient elevated temperature to cause pain and burns.The heating while charging was determined to be exacerbated by off-axis charging of shallow implanted ipgs and that all chargers were capable of elevated heating.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Manufacturer Narrative
(b)(4).Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Event Description
Device 1 of 3 reference mfr.Report#: 1627487-2015-05107 reference mfr.Report#: 1627487-2015-05184 follow-up revealed the replacement charging system did not provide resolution to the patient's pocket heating issue.In turn, another replacement charging system will be sent to the patient.
 
Manufacturer Narrative
Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOBILE CHARGING SYSTEM
Type of Device
SCS CHARGING SYSTEM
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key4582074
MDR Text Key5339808
Report Number1627487-2015-05106
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup,Followup
Report Date 03/11/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/09/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/28/2014
Device Model Number3711
Device Lot Number4108746
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received03/11/2015
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/07/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487-07262012-001-C
Patient Sequence Number1
Patient Outcome(s) Other;
-
-