EDWARDS LIFESCIENCES PR SWAN-GANZ CCOMBO V CCO/SVO2/CEDV THERMODILUTION CATHETER WITH AMC THROMBOSHIELD; CATHETER, OXIMETER, FIBEROPTIC Back to Search Results
Model Number 774HF75
Device Problem Material Rupture (1546)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/06/2015
Event Type  malfunction  
Event Description
It was reported that the balloon burst before use.No patient complications were reported.
 
Manufacturer Narrative
We received one catheter with an attached monoject 3ml syringe with 1.5 ml limited volume for examination.The balloon had a rupture in the central area of the balloon.Edges of latex did not appear to match up at the ruptured site, indicating a gap of missing latex.All other through lumens were patent without any leakage or occlusion.No other visual damage, contamination, or other abnormalities were found on the catheter and syringe.Lot number was not provided, therefore review of the manufacturing records could not be completed.An investigation has been initiated to consider any potential manufacturing factors that may have contributed to this complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ CCOMBO V CCO/SVO2/CEDV THERMODILUTION CATHETER WITH AMC THROMBOSHIELD
Type of Device
CATHETER, OXIMETER, FIBEROPTIC
Manufacturer (Section D)
EDWARDS LIFESCIENCES PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer (Section G)
EDWARDS LIFESCIENCES PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer Contact
lynn selawski
1 edwards way
irvine, CA 92614
9497564386
MDR Report Key4582166
MDR Text Key5341944
Report Number2015691-2015-00548
Device Sequence Number1
Product Code DQE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K040287
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 01/22/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number774HF75
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/19/2015
Initial Date Manufacturer Received 01/22/2015
Initial Date FDA Received03/09/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-