• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MED REL MEDTRONIC PUERTO RICO SURESCAN; STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MED REL MEDTRONIC PUERTO RICO SURESCAN; STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF) Back to Search Results
Model Number 97702
Device Problems High impedance (1291); Therapy Delivered to Incorrect Body Area (1508); Device Operates Differently Than Expected (2913)
Patient Problems Undesired Nerve Stimulation (1980); Pain (1994); Therapeutic Response, Decreased (2271)
Event Type  malfunction  
Event Description
It was reported that the patient had a loss of stimulation and therapeutic effect.The patient still felt stimulation but it was no longer effective to alleviate his pain.It was also noted that stimulation was in the wrong location.Upon interrogation, all electrodes had impedances of greater than 10,000 ohms.The patient had a limb fall on him that landed near the generator site a few months ago.It was unknown if this had anything to do with the impedance issues.X-rays were ordered to check lead placement.The x-rays were done on a walk in basis and it was unknown when these were done.The patient had a follow up appointment scheduled for (b)(6) 2015 to discuss the x-ray results.A possible revision was discussed but not planned.It was also clarified that the impedance issue was only on the 0 electrode when paired with all other electrodes.Impedances were not high when any of the electrodes were paired with any electrode other than 0.The 0 electrode was not used in any of the programming.Additional information regarding further troubleshooting, results of the xray and outcome were requested.If received, a follow up report will be sent.
 
Manufacturer Narrative
Concomitant medical products: product id: 977a260, serial# (b)(4), implanted: (b)(6) 2013, product type: lead.Product id: 977a260, serial# (b)(4), implanted: (b)(6) 2013, product type: lead.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESCAN
Type of Device
STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF)
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
ceiba norte industrial park, r
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC NEUROMODULATION
7000 central avenue ne rcw215
minneapolis MN 55432
Manufacturer Contact
diane wolf
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263987
MDR Report Key4600412
MDR Text Key5631869
Report Number3004209178-2015-04721
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 02/19/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/13/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/14/2014
Device Model Number97702
Device Catalogue Number97702
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received02/19/2015
Date Device Manufactured06/25/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age00073 YR
-
-