Catalog Number 2200261 |
Device Problems
Incorrect, Inadequate or Imprecise Result or Readings (1535); Incorrect Or Inadequate Test Results (2456); Output Problem (3005)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692); No Code Available (3191)
|
Event Date 02/13/2015 |
Event Type
malfunction
|
Event Description
|
On (b)(6) 2015, a laboratory personnel from (b)(4) indicated that a physician had questioned a patient's a1c result obtained on the d-10 hemoglobin testing system.(b)(4) confirmed there were no indications from the physician of any change of treatment or harm done to any patients based on the erroneous a1c result.(b)(4) stopped running the d-10 instrument on (b)(6) 2015.(b)(4) personnel said that they did a comparison for patient samples for the past 3 days (11, 12, 13) with results obtained from another lab.All patient results that were over 10% discrepancy tolerance were reported to the physician.(b)(4) personnel reported that 4 patients exceeded the 10% tolerance.The rest of the results matched the results they had obtained in their own lab.(b)(4) also confirmed that no patient samples were run since they reported the issue on (b)(6) 2015 until after the issue was resolved on (b)(6) 2015.
|
|
Manufacturer Narrative
|
An investigation is underway to determine root cause for the incident.
|
|
Manufacturer Narrative
|
A field service engineer was able to resolve the issue by replacing the detector.The detector has not been returned for analysis.A root cause has not been confirmed.Note: catalog number has been changed to reflect a detector part number, rather than that the overall testing system number.
|
|
Search Alerts/Recalls
|