• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. STRAIGHT SUCTION, AXIEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. STRAIGHT SUCTION, AXIEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number 9733449
Device Problems Bent (1059); Device Issue (2379); Material Deformation (2976)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/17/2015
Event Type  malfunction  
Manufacturer Narrative
Device manufacturing date now provided.
 
Event Description
A medtronic technical services representative reported that, while in a transsphenoidal procedure, the site's straight suction instrument was damaged.This occurred after initially verifying the instrument and navigating for two hours without issue.At this point, the surgeon was navigating in the sphenoid when alleging that the straight suction was inaccurate by 1 centimeter.In trouble-shooting, the patient was re-registered using tracer, however, the straight suction instrument could not be verified.The medtronic representative noted that the site was grabbing the instrument from the suction tubing instead of from the handle.The site was instructed to change the suction on the instrument by grabbing the handle, not the shaft and tubing, to avoid bending of the instrument.A medtronic ear, nose & throat (ent) representative brought a different straight suction in to be used for the remainder of the procedure.The surgeon continued and completed the procedure with the use of the navigation system.There was no impact on patient outcome.
 
Manufacturer Narrative
Return requested.Replacement straight suction shipped to site 02/17/2015.No parts have been received by manufacturer for analysis.Part not returned to manufacturer.
 
Manufacturer Narrative
Correction: on 13-oct-2015, it was noticed that a previous mdr submission contained incorrect information with regards to the common device name, product code and/or pma/510(k).This mdr is being submitted to correct this information.There is no new information to change the patient information, event description or manufacturer narrative that was previously reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRAIGHT SUCTION, AXIEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC.
826 coal creek circle
louisville CO 80027 9710
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC.
826 coal creek circle
louisville CO 80027 9710
Manufacturer Contact
kristianna bilan
826 coal creek circle
louisville, CO 80027-9710
7208902338
MDR Report Key4610536
MDR Text Key13190437
Report Number1723170-2015-00330
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative,company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup,Followup
Report Date 11/10/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Catalogue Number9733449
Device Lot Number25477
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/08/2015
Initial Date FDA Received03/17/2015
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received04/13/2015
11/11/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/12/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Weight82
-
-