Model Number 3688
Device Problems Premature End-of-Life Indicator (1480); Communication or Transmission Problem (2896); Unintended Movement (3026)
Patient Problem Inadequate Pain Relief (2388)
Event Type  Injury  
Manufacturer Narrative
Evaluation results: the device was subjected to a functional test on automated test equipment (ate).This tester verifies the electrical performance of the control/communication circuitry, and output signal integrity.All portions of ate testing passed.The pulse load test confirmed the device battery was depleted to the end of life.The programming history of the device was not provided, therefore an estimate of longevity could not be performed.The device performed in specification, and the root cause of the depleted battery could not be determined.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Event Description
It was reported the patient experienced difficulty with the ipg communicating with the programmer due to the ipg being tilted in the pocket.It was also reported the patient lost stimulation after receiving a "low battery" message.As a result, the patient's ipg was explanted and replaced.Surgical intervention resolved the patient's issues.The event date(s) related to the patient's issues are unknown.
 
Manufacturer Narrative
(b)(4).Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EON C 16-CHANNEL IPG
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key4614758
MDR Text Key13400580
Report Number1627487-2015-05136
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/28/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/28/2014
Device Model Number3688
Device Lot Number3651189
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/03/2015
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/28/2015
Initial Date FDA Received03/18/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/23/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/23/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MODEL: 3228, SCS LEAD
Patient Outcome(s) Other;
-
-