The following description of the event was documented by a carefusion technical support specialist in response to an email from the distributor in (b)(6)."oxygen concentration readings failed pre-delivery inspection out of spec.Out of box failure 21:23.6, 30:29.1, 60:56.1, 90:83.2, 100.97.2".
|
(b)(4).Carefusion issued a return goods authorization (rga) number for the return of the alleged faulty device for evaluation.As of the march 17, 2015 the alleged faulty device has not been received.Once the alleged faulty device is received and evaluated, a follow-up medwatch report will be submitted.
|
Corrected serial # to (b)(4).Failure analysis: microblender pn: 03920a, sn: (b)(4) was received into the carefusion failure analysis lab for investigation.The carefusion failure analysis lab technician examined the microblender and found that it is out of calibration.The microblender was tested on test station # 10464 per drs 90290 and found that it is out of calibration at 30%, 60% and 90% o2.At end point calibration sequence # 1, 21%, spec.Is 20.9 to 22.9% reads 21.8%.At set point calibration sequence # 3, 60%, spec.Is 58.0 to 62.0% reads 54.8%.At set point calibration sequence # 4, 90%, spec.Is 88.0 to 92.0% reads 86.1%.At set point calibration sequence # 5, 100%, spec.Is 98.0 to 100.0% reads 98.7%.At set point calibration sequence # 7, 60%, spec.Is 58.0 to 62.0% reads 54.6%.At set point calibration sequence # 8, 60%, spec.Is 57.0 to 63.0% reads 54.9%.At set point calibration sequence # 9, 60%, spec.Is 57.0 to 63.0% reads 53.5%.Additional issues related to complaint; at set point calibration sequence # 2, 30%, spec.Is 28.0 to 32.0% reads 25.0%.The microblender was recalibrated by adjusting the valve seat and the control knob.Finding/root-cause: the microblender was out of calibration at 60% and 90% but was not out of calibration at 21% and 100% as alleged, but was out of calibration at 30%.
|