The following description of the event was documented by a carefusion technical support specialist in response to an email from the distributor in australia."oxygen concentration readings failed pre-delivery inspection out of spec.Out of box failure 21:30.3, 30:32.5, 60:56.1, 90:83.2, 100:97.2.".
|
(b)(4).Carefusion issued a return goods authorization (rga) number for the return of the alleged faulty device for evaluation.As of the march 17, 2015 the alleged faulty device has not been received.Once the alleged faulty device is received and evaluated, a follow-up medwatch report will be submitted.
|
(b)(4).Failure analysis: microblender pn: 03920a, sn: (b)(4) was received into the carefusion failure analysis lab for investigation.The carefusion failure analysis lab technician examined the microblender and found that it is out of calibration.The microblender was tested on test station (b)(4) and found that it is out of calibration at 21% and 30% o2.At end point calibration sequence # 1, 21%, spec.Is 20.9 to 22.9% reads 26.7%.At set point calibration sequence # 2, 30%, spec.Is 28.0 to 32.0% reads 27.8%.At set point calibration sequence # 3, 60%, spec.Is 58.0 to 62.0% reads 58.0%.At set point calibration sequence # 4, 90%, spec.Is 88.0 to 92.0% reads 91.1%.At set point calibration sequence # 5, 100%, spec.Is 98.0 to 100.0% reads 99.6%.At set point calibration sequence # 7, 60%, spec.Is 58.0 to 62.0% reads 58.8%.At set point calibration sequence # 8, 60%, spec.Is 57.0 to 63.0% reads 59.2%.At set point calibration sequence # 9, 60%, spec.Is 57.0 to 63.0% reads 58.7%.The microblender was recalibrated by adjusting the valve seat and the control knob.However, it was found that the best calibration yielded readings of at 21%; 22.7, 30%; 29.2, 60%; 60.3, 90%; 89.1, 100%; 100.0.The poppet pn: 03929 was replaced with a known good poppet and now the microblender can be calibrated as follows 21%; 21.3, 30%; 29.9, 60%; 59.5, 90%; 89.5, 100%; 99.6.Duplicated, the microblender is out of calibration at 21% and 30% but did not duplicate microblender out of calibration at 60%, 90% and 100% complaint allegation.Finding/root-cause: the microblender was out of calibration due to a faulty poppet pn: 03929.
|