ETHICON INC. GYNECARE TVT OBTURATOR SYSTEM; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC Back to Search Results
Catalog Number 810081L
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/21/2015
Event Type  malfunction  
Event Description
It was reported that a patient underwent a sling procedure due to incontinence on (b)(6) 2015.During the procedure, the plastic tube was damaged.Another like device was used with no adverse patient consequences.
 
Manufacturer Narrative
(b)(4).Conclusion: the actual device was returned for evaluation.One needle is broken.The break is located at the level of the introduction hole.Presence of clamp marks were found on the broken needle in two different locations.In addition, a review of the batch manufacturing records was conducted and the batch met all finished goods release criteria.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GYNECARE TVT OBTURATOR SYSTEM
Type of Device
MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876 015
Manufacturer (Section G)
ETHICON SARL-NEUCHATEL
puits-godet 20 2000 neuchatel
neuchatel
SZ  
Manufacturer Contact
guillermo villa
route 22 west po box 151
somerville, NJ 08876
9082180707
MDR Report Key4627997
MDR Text Key5660555
Report Number2210968-2015-03450
Device Sequence Number1
Product Code OTN
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K033568
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 01/23/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/24/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2015
Device Catalogue Number810081L
Device Lot Number3805343
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/23/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received03/24/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-