• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION LAMITRODE S-8 LEAD KIT, 60CM LENGTH; SCS LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION LAMITRODE S-8 LEAD KIT, 60CM LENGTH; SCS LEAD Back to Search Results
Model Number 3286
Device Problems Fracture (1260); High impedance (1291); Therapy Delivered to Incorrect Body Area (1508)
Patient Problem Therapeutic Effects, Unexpected (2099)
Event Date 03/04/2015
Event Type  Injury  
Event Description
Device 1 of 2.Reference mfr.Report#: 1627487-2015-05152.It was reported the patient is receiving stimulation in an unintended area.Reprogramming attempts were unable to provide resolution.An impedance check revealed high impedance.X-rays were taken and the doctor believes one of the leads is fractured due to the results.As a result, the patient will undergo surgical intervention.Both of the patient's leads are being reported because it is unknown which lead is contributing to the patient's issue.
 
Manufacturer Narrative
Udi (di): (b)(4).Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Event Description
Device 1 of 2.Reference mfr.Report#: 1627487-2015-05152.Follow-up revealed the patient's leads were explanted and replaced.Surgical intervention resolved the patient's issue.
 
Manufacturer Narrative
Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAMITRODE S-8 LEAD KIT, 60CM LENGTH
Type of Device
SCS LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key4628696
MDR Text Key21327120
Report Number1627487-2015-05151
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/26/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/28/2015
Device Model Number3286
Device Lot Number4000169
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/26/2015
Initial Date FDA Received03/24/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/21/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/21/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MODEL: 3788, SCS IPG; MODEL: 1192(X2), SCS ANCHORS
Patient Outcome(s) Other;
Patient Age42 YR
-
-