Model Number 3286 |
Device Problems
Fracture (1260); High impedance (1291); Therapy Delivered to Incorrect Body Area (1508)
|
Patient Problem
Therapeutic Effects, Unexpected (2099)
|
Event Date 03/04/2015 |
Event Type
Injury
|
Event Description
|
Device 1 of 2.Reference mfr.Report#: 1627487-2015-05152.It was reported the patient is receiving stimulation in an unintended area.Reprogramming attempts were unable to provide resolution.An impedance check revealed high impedance.X-rays were taken and the doctor believes one of the leads is fractured due to the results.As a result, the patient will undergo surgical intervention.Both of the patient's leads are being reported because it is unknown which lead is contributing to the patient's issue.
|
|
Manufacturer Narrative
|
Udi (di): (b)(4).Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
|
|
Event Description
|
Device 1 of 2.Reference mfr.Report#: 1627487-2015-05152.Follow-up revealed the patient's leads were explanted and replaced.Surgical intervention resolved the patient's issue.
|
|
Manufacturer Narrative
|
Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
|
|