Model Number 0295
Device Problems Over-Sensing (1438); Pocket Stimulation (1463); Device Sensing Problem (2917); Human-Device Interface Problem (2949)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/17/2015
Event Type  Injury  
Event Description
Boston scientific received information that the patient with this right ventricular (rv) lead experience muscle stimulation.The lead also displayed noise and oversensing.It was suspected that the lead had perforate the patient's ventricle.The patient underwent a surgical revision procedure where the lead was surgically abandoned and replaced.There were no additional adverse patient effects reported.
 
Manufacturer Narrative
(b)(4).As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOTAK RELIANCE
Manufacturer (Section D)
CPI - DEL CARIBE
guidant puerto rico b. v.
dorado PR
Manufacturer (Section G)
CPI - DEL CARIBE
guidant puerto rico b. v.
dorado PR
Manufacturer Contact
sonali vasekar
4100 hamline ave. n
st. paul, MN 55112
6515824786
MDR Report Key4654679
MDR Text Key5681813
Report Number2124215-2015-02935
Device Sequence Number1
Product Code NVY
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/17/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/02/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/30/2014
Device Model Number0295
Other Device ID NumberRELIANCE 4-SITE
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Manufacturer Received02/17/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/30/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
N140; 4543; 4470; 0295
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age76 YR
-
-