Catalog Number 0250070520 |
Device Problem
Device Packaging Compromised (2916)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/12/2015 |
Event Type
malfunction
|
Event Description
|
It was reported that there was a hair inside the sterile pack.
|
|
Manufacturer Narrative
|
Additional information will be provided once the investigation has been completed.
|
|
Manufacturer Narrative
|
The foreign material failure mode was confirmed on the unit returned.A small strand of hair was found inside the blister package.Most probable root causes are: manufacturing/assembly error, incorrect or inadequate packaging.As per the evidence at hand root cause is a manufacturing error.This was deduced due to current risk documents and investigation.As per current risk documents, the event is within predicted acceptable occurrence and severity thresholds.The risk is within the green low acceptable level.There are mitigating procedures in place.No further actions are deemed necessary at this time the product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.
|
|
Event Description
|
It was reported that there was a hair inside the sterile pack.
|
|