ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) PROTEGE IPG, 16-CHANNEL RECHARGEABLE; SCS IPG Back to Search Results
Model Number 3789
Device Problems Charging Problem (2892); Unintended Movement (3026)
Patient Problem Weight Changes (2607)
Event Date 02/20/2015
Event Type  Injury  
Event Description
It was reported the patient's scs ipg charger displays the slow-flashing green led, indicating ipg-charging at a reduced efficiency.The patient stated the most recent recharge required about 3.5 hours to get to 75% full.A sjm representative met with the patient and confirmed the patient's ipg appears titled in the pocket.Also, the patient has lost approximately 20 pounds since the implant.Follow-up identified surgical intervention was undertaken on (b)(6) 2015 to revise the patient's ipg pocket.The physician made a new pocket to the side of the original pocket which was more superficial and level.
 
Manufacturer Narrative
(b)(4).Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROTEGE IPG, 16-CHANNEL RECHARGEABLE
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
MDR Report Key4666133
MDR Text Key5728015
Report Number1627487-2015-01139
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 03/16/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Other
Device Expiration Date09/30/2016
Device Model Number3789
Device Lot Number4793565
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/16/2015
Initial Date FDA Received04/07/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/18/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MODEL 3228, SCS LEAD
Patient Outcome(s) Other;
Patient Age40 YR
-
-