It was reported the patient's scs ipg charger displays the slow-flashing green led, indicating ipg-charging at a reduced efficiency.The patient stated the most recent recharge required about 3.5 hours to get to 75% full.A sjm representative met with the patient and confirmed the patient's ipg appears titled in the pocket.Also, the patient has lost approximately 20 pounds since the implant.Follow-up identified surgical intervention was undertaken on (b)(6) 2015 to revise the patient's ipg pocket.The physician made a new pocket to the side of the original pocket which was more superficial and level.
|