Model Number CD3365-40Q
Device Problem Difficult to Interrogate (1331)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/13/2015
Event Type  malfunction  
Event Description
It was reported that the patient presented in the operating room for new system implant.Device was initially interrogated successfully, however, rf telemetry was intermittent and the programmer was not able to locate the device.Programmer was rebooted multiple times without resolving the issue.The device was not implanted.Patient condition post procedure was good.
 
Manufacturer Narrative
All information provided by manufacturer, no medwatch form was received.(b)(4).
 
Manufacturer Narrative
The reported rf telemetry anomaly was not confirmed in the laboratory.The device was tested on the bench and no anomaly was found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA CRT-D, DF-4 CONNECTOR
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC., CRMD
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC., CRMD
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sharon mathew
645 almanor avenue
sunnyvale, CA 94085
4085226327
MDR Report Key4670034
MDR Text Key5728146
Report Number2938836-2015-03365
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P910023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2017
Device Model NumberCD3365-40Q
Device Catalogue NumberSMTFY999
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/30/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/13/2015
Initial Date FDA Received04/08/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/24/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/27/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-