Model Number DTBB1D1
Device Problems Inappropriate/Inadequate Shock/Stimulation (1574); Invalid Sensing (2293)
Patient Problem Shock (2072)
Event Date 01/07/2015
Event Type  Injury  
Event Description
It was reported that the patient received two shocks.There was anti-tachycardia pacing (atp) during charging and delivery occurred.The rhythm was described as a "dirty break" that was redetected with ventricular fibrillation (vf) therapy with the second committed and delivered.The clinician deemed the second therapy inappropriate.The implantable cardioverter defibrillator (icd) remains in use.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.Concomitant medical products: 6932-65, lead, implanted: (b)(6) 1997; 419688, lead, implanted: (b)(6) 2014; 5076-52, lead, implanted: (b)(6) 2014.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIVA S
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM HEART FAILURE
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key4674671
MDR Text Key5642195
Report Number3004209178-2015-06298
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 01/08/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/28/2015
Device Model NumberDTBB1D1
Device Catalogue NumberDTBB1D1
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/08/2015
Initial Date FDA Received04/09/2015
Date Device Manufactured02/07/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
693352 LEAD
Patient Outcome(s) Life Threatening;
Patient Age00066 YR
-
-