• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IPG MFG SWITZERLAND EGIDA CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IPG MFG SWITZERLAND EGIDA CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO Back to Search Results
Model Number D394TRG
Device Problems Inappropriate/Inadequate Shock/Stimulation (1574); Invalid Sensing (2293)
Patient Problem Shock (2072)
Event Date 01/16/2015
Event Type  Injury  
Event Description
It was reported that the patient received an inappropriate therapy.It was noted there was an undersensing issue during the ventricular fibrillation (vf) episode.The device was reprogrammed.The device and lead remain in use.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
This model number is not approved for distribution in the united states; however, it is same/similar to a device marketed in the u.S.This event occurred outside the us and patient information is not generally available due to confidentiality concerns.(b)(4).
 
Manufacturer Narrative
Product event summary: the device was not returned for analysis.However, performance data collected from the device was received and analyzed.Analysis of the device memory indicated rv (right ventricular) undersensing information.Intervals of undersensing noted in vf episode.Device appropriately detected and delivered by therapy.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EGIDA CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
IPG MFG SWITZERLAND
route du molliau 31
tolochenaz 1131
CH  1131
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM HEART FAILURE
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key4675047
MDR Text Key5686372
Report Number9614453-2015-00737
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P010031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,company representati
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 01/16/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/14/2014
Device Model NumberD394TRG
Device Catalogue NumberD394TRG
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/17/2015
Initial Date FDA Received04/09/2015
Supplement Dates Manufacturer ReceivedNot provided
04/17/2015
Supplement Dates FDA Received06/10/2015
09/18/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/16/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
693565 LEAD, 419488 LEAD
Patient Outcome(s) Life Threatening; Required Intervention;
Patient Age00075 YR
-
-