DEPUY ORTHOPAEDICS, INC.1818910 SELF CENT HIP 41X28 RUST; HIP OTHER IMPLANT Back to Search Results
Catalog Number 103541000
Device Problems Disassembly (1168); Fracture (1260)
Patient Problem No Information (3190)
Event Date 04/06/2015
Event Type  Injury  
Event Description
Patient was revised to address a broken bipolar liner, which caused the bipolar head to disassociate.
 
Manufacturer Narrative
This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).
 
Manufacturer Narrative
Examination of the reported device was not possible as it was not returned.A search of the complaints databases finds no other reports against the product and lot code combination since its release to distribution.An x-ray was provided for examination confirming the reported implant disassociation.The femur is secured with a trauma plate and several control cables, one of which is broken, and the bipolar head has rotated into a vertical position.No medical records or additional event information was provided to customer quality.The investigation can draw no conclusion with the information provided.Based on the inability to determine root cause, the need for corrective action has not been indicated.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SELF CENT HIP 41X28 RUST
Type of Device
HIP OTHER IMPLANT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC.1818910
700 orthopaedic drive
warsaw IN 46582
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC.1818910
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
steve dowell
700 orthopaedic drive
warsaw, IN 46581
5743714918
MDR Report Key4690381
MDR Text Key5711900
Report Number1818910-2015-18465
Device Sequence Number1
Product Code KWY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK812672
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor,distributor,health prof
Reporter Occupation Other
Type of Report Initial
Report Date 04/06/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date11/22/2005
Device Catalogue Number103541000
Device Lot NumberU6BAW1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/06/2015
Initial Date FDA Received04/14/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/22/2000
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age54 YR
-
-