The device history record was reviewed indicating that product was released accomplishing all quality standard requirements.There were no non-conforming issues reported during the production of this product for a similar condition as reported in this complaint.The complaint sample was not returned to the manufacturing site for review.Without the sample, it is not possible to determine a confirmed root cause of this issue.Should the sample be returned in the future, this complaint will be re-opened for further investigation.A corrective action is not warranted at this time.It must be noted that in-process controls (such as personnel training, incoming quality acceptance testing for raw material, in process visual inspection and visual acceptance sampling) are in place to prevent non-conforming product from leaving the manufacturing operations.As per procedure, manufacturing performs 100% leak testing at the final stage of production, which would identify this issue in the catheter assembly.This complaint will be used for tracking and trending purposes.
|