COVIDIEN PALINDROME 23/40 KIT W/ SLOT; DIALYSIS CATHETER Back to Search Results
Model Number 8888145015
Device Problems Air Leak (1008); Crack (1135)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/13/2015
Event Type  malfunction  
Event Description
It was reported to covidien on (b)(6) 2015 that a customer had an issue with a dialysis catheter.The customer stated that there was a crack in the joint of the catheter near the skin after the product was placed and used for 11 months.There were many bubbles in the blood while conducting dialysis.The doctor immediately stopped using it and replaced the catheter for the patient.The patient was involved without injury.The patient is in good condition.
 
Manufacturer Narrative
An investigation is currently underway, upon completion the results will be forwarded.
 
Manufacturer Narrative
The device history record was reviewed indicating that product was released accomplishing all quality standard requirements.There were no non-conforming issues reported during the production of this product for a similar condition as reported in this complaint.The complaint sample was not returned to the manufacturing site for review.Without the sample, it is not possible to determine a confirmed root cause of this issue.Should the sample be returned in the future, this complaint will be re-opened for further investigation.A corrective action is not warranted at this time.It must be noted that in-process controls (such as personnel training, incoming quality acceptance testing for raw material, in process visual inspection and visual acceptance sampling) are in place to prevent non-conforming product from leaving the manufacturing operations.As per procedure, manufacturing performs 100% leak testing at the final stage of production, which would identify this issue in the catheter assembly.This complaint will be used for tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PALINDROME 23/40 KIT W/ SLOT
Type of Device
DIALYSIS CATHETER
Manufacturer (Section D)
COVIDIEN
covidien manufacturing solulfons sa
edificio 820 calle #2 zona franca coyol
alajuela
CS 
Manufacturer (Section G)
COVIDIEN
covidien manufacturing solulfons sa
edificio 820 calle #2 zona franca coyol
alajuela
CS  
Manufacturer Contact
thom mcnamara
15 hampshire street
mansfield, MA 02048
5084524811
MDR Report Key4704209
MDR Text Key5716494
Report Number3009211636-2015-00207
Device Sequence Number1
Product Code MSD
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/14/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/17/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8888145015
Device Catalogue Number8888145015
Device Lot Number222926X
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/15/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/05/2012
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
-
-