Model Number 1900
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/19/2015
Event Type  malfunction  
Event Description
Hill-rom received a report from the account stating the left hand cpr leer was loose.The bed was located at he account in room 516.There was no patient /user injury reported.This report was filed in our complaint handling system as complaint #(b)(4).
 
Manufacturer Narrative
The hill-rom technician found a broken roll pin on cpr lever causing the cpr not to work.A search of the hill-rom maintenance records showed hill-rom performed preventative maintenance on this bed in 2014.It is unknown if the facility performed any other preventative maintenance on this bed.The technician replaced the broken roll pin on cpr lever to resolve the issue.Based on this information, no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOTALCARE BED
Type of Device
A/C POWERED ADJUSTABLE HOSPITAL BED
Manufacturer (Section D)
HILL-ROM, INC.
batesville IN 47006
Manufacturer Contact
john cummings
1069 state route 46 east
batesville, IN 47006
8129312869
MDR Report Key4713851
MDR Text Key20365673
Report Number1824206-2015-00502
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122473
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Unknown
Type of Report Initial
Report Date 03/19/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number1900
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/19/2015
Initial Date FDA Received04/15/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-