Catalog Number 201.928
Device Problem Material Fragmentation (1261)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 04/04/2015
Event Type  Injury  
Event Description
Device report synthes europe reported an event in (b)(6) as follows: it was reported that when the surgeon was implanting the screw the screw head broke.The screw shaft remains in the patient¿s bone.No surgical delay or patient harm were reported.This report is 1 of 1 for (b)(4).
 
Manufacturer Narrative
Device was used for treatment, not diagnosis.Reporter phone number: (b)(6).Without a lot number the device history records review could not be completed.The investigation could not be completed; no conclusion could be drawn, as no product was received.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
A product investigation was completed: our investigation has shown that the screw indeed is broken off as complained.Unfortunately we are not able to determine the exact cause which has led to this breakage because no further information has made been available.The device history record could not be reviewed as the lot number is unknown.It is likely that too much mechanical force had been applied while placing.The type and extent of damage incurred indicate that this complaint was caused by wrong handling.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
The subject device has been received and is currently in the evaluation process.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Device broke during insertion; device was not considered implanted/explanted.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.0MM IMF SCREW SELF-DRILLING 8MM
Type of Device
SCREW, FIXATION, INTRAOSSEOUS
Manufacturer (Section D)
SYNTHES (USA)
1302 wrights lane east
west chester PA 19380
Manufacturer Contact
linda plews
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key4721325
MDR Text Key20753748
Report Number2520274-2015-13334
Device Sequence Number1
Product Code DZL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK010527
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 04/07/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number201.928
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/02/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/30/2015
Initial Date FDA Received04/22/2015
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Not provided
Supplement Dates FDA Received05/29/2015
06/16/2015
07/06/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;