Model Number 60-02-15
Device Problem Other (for use when an appropriate device code cannot be identified) (2203)
Patient Problem No Patient involvement (2648)
Event Date 03/30/2015
Event Type  Other  
Event Description
Sorin group (b)(4) received a report that the centrifugal pump system did not detect flow during set up.There was no patient involvement.
 
Manufacturer Narrative
Sorin group (b)(4) manufactures the sorin centrifugal pump system with tubing clamp.The incident occurred in (b)(6).This medwatch report is filed on behalf of sorin group (b)(4).Sorin group received a report that the centrifugal pump system did not detect flow during set up.There was no patient involvement.The investigation is ongoing.A follow-up report will be sent when the investigation is complete.
 
Manufacturer Narrative
A sorin group representative removed the flow sensor and replaced it with a new sensor.After installation the new sensor was tested and then the equipment was returned to service.The customer retained the unit.Without the unit to evaluate, no root cause could be identified and no corrective action could be determined.No nonconformities were noted during the device history record review regarding this issue.The issue will be monitored for trends and if identified, corrections will be recommended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SORIN CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND
lindbergh strasse 25
munich D 809 39
GM  D 80939
Manufacturer (Section G)
SORIN GROUP DEUTSCHLAND
lindbergh strasse 25
munich D 809 39
GM   D 80939
Manufacturer Contact
cheri voorhees, mgr., qa
14401 w. 65th way
arvada, CO 80004
3034676527
MDR Report Key4724823
MDR Text Key5747121
Report Number9611109-2015-00118
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K032213
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,health professional
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date 03/30/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/20/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-02-15
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/30/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/01/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-