Boston scientific received information from the patient's family that this patient died when the generator was replaced.Further review of the records, shows that the patient died approximately five days after the generator replacement, and according to the on-line obituary the death occurred in the hospital.Further information has been requested from the field.Additional information was received from the local field representative.At the time of the generator change-out the patient had complications attributed to anesthesia and went into respiratory arrest.A code was called.There were no device related issues, but the lower pacing rate limit was increased to provide additional support at the time.The patient was transferred in unstable condition to the intensive care unit.The field representative was never contacted for any additional device evaluations.
|