• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HILL-ROM, INC. AFFINITY 4 BED; BIRTHING BED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM, INC. AFFINITY 4 BED; BIRTHING BED Back to Search Results
Model Number 3700
Device Problem Self-Activation or Keying (1557)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/25/2015
Event Type  malfunction  
Event Description
Hill-rom received a report from the account stating the head section would self-run when the bed is plugged in.The bed was located in the maintenance shop at the account.There was no pt/user injury reported.This report was filed in our complaint handling system as complaint # (b)(4).
 
Manufacturer Narrative
Hill-rom technical support suggested isolating the left side rail.Per the hill-rom service manual, the affinity three birthing bed and affinity four birthing beds require an effective maintenance program.We recommend that you perform semiannual preventive maintenance (pm) along with a quarterly battery check and testing for joint commission on accreditation of healthcare organizations (jcaho).Pm and testing not only meet jcaho requirements but will help ensure a long, operative life for the bed.Pm will minimize downtime due to excessive wear.Check the switches in the side rails to ensure they are functioning correctly.Also check for intermittent operation.A search of the hill-rom maintenance records did not show hill-rom performed any preventative maintenance on this bed.It is unk if the facility performs preventative maintenance on their beds.No further info is available on the repair of the bed at this time.The investigation is ongoing, however if any additional relevant info is identified following completion of the investigation, the additional relevant info will be submitted in a supplemental report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFFINITY 4 BED
Type of Device
BIRTHING BED
Manufacturer (Section D)
HILL-ROM, INC.
batesville IN
Manufacturer Contact
tony werner
1069 state route 46 east
batesville, IN 47006
8129312359
MDR Report Key4731058
MDR Text Key5744277
Report Number1824206-2015-00521
Device Sequence Number1
Product Code KNC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K915779
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Unknown
Type of Report Initial
Report Date 03/25/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number3700
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/25/2015
Initial Date FDA Received04/21/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-