• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, INC. SYNCARDIA FREEDOM DRIVER; CIRCULATORY ASSIST DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, INC. SYNCARDIA FREEDOM DRIVER; CIRCULATORY ASSIST DEVICE Back to Search Results
Catalog Number 595000-001
Device Problems Difficult to Insert (1316); Difficult to Remove (1528)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/03/2015
Event Type  malfunction  
Event Description
The customer reported that it was difficult to insert and extract the freedom onboard batteries from the freedom driver's battery wells.The patient was subsequently switched to the backup driver.There was no reported adverse patient impact.This alleged failure mode poses a low risk to the patient because although the onboard batteries were difficult to insert and extract from the driver battery wells, the freedom driver continued to perform its life-sustaining functions.In addition, the freedom driver also has a redundant power source of external wall power.The freedom driver will be returned to syncardia for evaluation.The results of the investigation will be provided in a supplemental mdr.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
CIRCULATORY ASSIST DEVICE
Manufacturer (Section D)
SYNCARDIA SYSTEMS, INC.
tucson AZ
Manufacturer Contact
carole marcot,esq.
1992 e. silverlake rd.
tucson, AZ 85713-CE 2
5205451234
MDR Report Key4737350
MDR Text Key5826783
Report Number3003761017-2015-00132
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Distributor
Reporter Occupation Other Health Care Professional
Remedial Action Replace
Type of Report Initial
Report Date 04/03/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/24/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number595000-001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/03/2015
Was Device Evaluated by Manufacturer? No
Date Device Manufactured12/01/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age45 YR
-
-