MAUDE Adverse Event Report: CODMAN AND SHURTLEFF, INC ORBIT GALAXY G2 MICROCOIL DELIVERY SYSTEM; ARTIFICIAL EMBOLIZATION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CODMAN AND SHURTLEFF, INC ORBIT GALAXY G2 MICROCOIL DELIVERY SYSTEM; ARTIFICIAL EMBOLIZATION DEVICE Back to Search Results
Catalog Number 641CX2505
Device Problem Difficult To Position (1467)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/06/2015
Event Type  malfunction  
Event Description
The contact at the hospital reported that the doctor tried to deploy the presidio coil (details unknown) and orbit galaxy g2 (641cx2505/c15306) through the echelon microcatheter (details unknown) but the microcatheter faced stiff resistance and the coil did not break properly inside the aneurysm and after two loops, the catheter kicked back.The doctor used a similar sized target coil from stryker and it deployed very well.There was no reported patient impact or significant procedural delay.At the time of initial contact the orbit galaxy g2 was available for return.
 
Manufacturer Narrative
Concomitant products: presidio coil (details unknown); echelon microcatheter (details unknown).The product will be returned for analysis, however it has not been received.Additional information will be submitted within 30 days of receipt.No conclusions are made at this time.This report is related to mfr.Report # 2954740-2015-00100.
 
Manufacturer Narrative
The coil was returned undamaged.The coil¿s socket ring was pushed down inside the outer sheath (angle ring section).This caused a loss of concentricity between the distal tip of the device positioning unit (dpu) and the proximal end of the coil.This loss of concentricity can contribute to the coil¿s resistance inside the microcatheter.Located on the top proximal end of the resheathing tool in the open cutout section, the v notch has been severely fractured with the damaged edges raised above the surface plane.The locking mechanism has stretching and compression damage.No manufacturing defects were found.The most likely contributing factor to the coils resistance inside the microcatheter, the microcatheters kickback, and the coils positioning difficulty may have occurred when the microcoil system was first unlocked for use and the sheath was retracted straight back instead of up at a forty-five degree angle and then back.When the sheath was pulled straight back, the locking mechanism caught the inside of the v notch of the resheathing tool and became embedded.In addition, the locking mechanism may not have been fully disengaged off the core wire.The sheath also again caught the v notches extended edges which may have occurred later in the procedure at the target site.This produced a binding action between the device positioning unit (dpu), the sheath, and the coil.This binding action produced significant resistance which caused the coils resistance inside the microcatheter, pushed the coils socket ring down inside the outer sheath, pushed the microcatheter off the target site, and caused the positioning difficulty.For optimum product performance and to prevent potential complications, the instructions for use (ifu) recommends, ¿hold the introducer sheath (loosely-looped) in the left hand.Keeping the introducer tip near the re-sheathing tool, grasp the distal end of the re-sheathing tool between your left thumb and forefinger.Grasp the clear tab near the end of the introducer sheath body with the thumb and forefinger of your other hand.Gently pull the clear tab of the introducer sheath out and away from the re sheathing tool at a 45-degree angle to unlock the microcoil.Continue to pull the tab until an additional 0.5 to 1.0 inches (1.3 to 2.5 cm) of the translucent material is exposed.Gently fold the translucent tab towards the distal end, and firmly grasp the distal end of the re sheathing tool and the translucent tab between your thumb and forefinger, as shown in figure 3¿¿ caution: if unusual friction is noticed during advancement or retraction of the microcoil system, verify the locking mechanism, or clear tab is unlocked and pulled out from the resheathing tool approximately 1in.(2-3cm).¿ in addition, without the return of the unspecified echelon microcatheter used in the procedure, it cannot be determined if this component had any additional contributions to the complaint event.Dhr: a review of the manufacturing documentation associated with this lot presented no issues during the manufacturing or inspection process that can be related to the reported complaint.Based on the information and the damages to the device the failures reported by the customer are plausible, however procedural factors outlined in the ifu likely contributed to the event.Without return of concomitant products it cannot be determined if those components had any additional contributions to the complaint event.Additionally, review of this lot presented no issues during the manufacturing or inspection process that can be related to the reported complaint.Therefore, no corrective actions will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORBIT GALAXY G2 MICROCOIL DELIVERY SYSTEM
Type of Device
ARTIFICIAL EMBOLIZATION DEVICE
Manufacturer (Section D)
CODMAN AND SHURTLEFF, INC
325 paramount dr
raynham MA
Manufacturer Contact
duane durbin
821 fox lane
san jose, CA 95131
5088288310
MDR Report Key4737957
MDR Text Key5774667
Report Number2954740-2015-00100
Device Sequence Number1
Product Code HCG
Combination Product (y/n)N
PMA/PMN Number
K120686
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/11/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/27/2018
Device Catalogue Number641CX2505
Device Lot NumberC15306
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/02/2015
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/11/2015
Initial Date FDA Received04/29/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/11/2015
Date Device Manufactured02/28/2013
Is the Device Single Use? Yes
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-