• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. GYNECARE THERMACHOICE III UTERINE BALLOON THERAPY; DEVICE, THERMAL ABLATION, ENDOMETRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. GYNECARE THERMACHOICE III UTERINE BALLOON THERAPY; DEVICE, THERMAL ABLATION, ENDOMETRIAL Back to Search Results
Catalog Number TC003
Device Problem Balloon leak(s) (1052)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/06/2015
Event Type  malfunction  
Event Description
It was reported that the patient underwent thermal ablation on (b)(6) 2015.During the procedure, the balloon was tested and passed and when the balloon was inserted and therapy started, the pressure would not hold.When the surgeon removed the balloon, a leak was discovered.The procedure was completed with a like device.There were no patient consequences reported.
 
Manufacturer Narrative
(b)(4).Conclusion: to date, the device has not been returned.If the product is returned for evaluation, any further info derived from the evaluation will be submitted in a supplemental 3500a form.In addition, a review of the batch manufacturing records was conducted and the batch met all finished goods release criteria.
 
Manufacturer Narrative
It was reported that the patient is doing well.Conclusion: the actual device involved in this event was returned for evaluation.The evaluation found a pinhole on the balloon.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GYNECARE THERMACHOICE III UTERINE BALLOON THERAPY
Type of Device
DEVICE, THERMAL ABLATION, ENDOMETRIAL
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876 015
Manufacturer (Section G)
ETHICON INC.-JUAREZ
avenida de las torres 7125, co
l salvacar
ciudad juarez 3260 4
MX   32604
Manufacturer Contact
guillermo villa
route 22 west po box 151
somerville, NJ 08876
9082180707
MDR Report Key4742851
MDR Text Key5702830
Report Number2210968-2015-05144
Device Sequence Number1
Product Code MNB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/08/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/01/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/28/2016
Device Catalogue NumberTC003
Device Lot NumberHLMG09
Other Device ID Number10705031050730
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/05/2015
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received05/21/2015
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/27/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-