• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SORIN GROUP DEUTSCHLAND SORIN CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SORIN GROUP DEUTSCHLAND SORIN CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 60-02-15
Device Problem Pumping Stopped (1503)
Patient Problem No Patient involvement (2648)
Event Date 04/06/2015
Event Type  Other  
Event Description
Sorin group received a report that the centrifugal pump locked up during a setup.There was no pt involvement.
 
Manufacturer Narrative
Pt info was not provided.Sorin group (b)(4) manufactures the sorin centrifugal pump system with tubing clamp.The incident occurred in (b)(6).This medwatch report is filed on behalf of sorin group (b)(4).Sorin group received a report that the centrifugal pump locked up during a setup.There was no pt involvement.The investigation is ongoing.A follow-up report will be sent when the investigation is complete.See scanned page.
 
Manufacturer Narrative
Livanova (b)(4) manufactures the centrifugal pump system.The incident occurred in (b)(6).This medwatch report is being filed on behalf of livanova (b)(4).The control panel of the centrifugal pump system was returned to livanova (b)(4) for investigation.Visual inspection did not identify any abnormalities or defects.Functional testing and a hardware analysis were unable to reproduce the reported issue.As the issue could not be reproduced, a root cause was not determined.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SORIN CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich D 809 39
GM  D 80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key4747696
MDR Text Key5706472
Report Number9611109-2015-00130
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K032213
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,health professional
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date 05/09/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/29/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-02-15
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received04/10/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-