• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SORIN GROUP DEUTSCHLAND SORIN CENTRIFUGAL PUMP 5 (CP5); CARDIOPULMONARY BYPASS PUMP SPEED CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SORIN GROUP DEUTSCHLAND SORIN CENTRIFUGAL PUMP 5 (CP5); CARDIOPULMONARY BYPASS PUMP SPEED CONTROL Back to Search Results
Model Number 60-02-60
Device Problem Other (for use when an appropriate device code cannot be identified) (2203)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/31/2015
Event Type  malfunction  
Event Description
Sorin group received a report that the centrifugal pump momentarily turned off and then came back on during a procedure.There was no report of pt injury.
 
Manufacturer Narrative
Pt info was not provided.Sorin group (b)(4) manufactures the sorin centrifugal pump 5 (cp5).The incident occurred in (b)(6).This medwatch report is filed on behalf of sorin group (b)(4).Sorin group received a report that the centrifugal pump momentarily turned off and then came back on during a procedure.There was no report of pt injury.The investigation is ongoing.A follow-up report will be sent when the investigation is complete.See scanned page.
 
Manufacturer Narrative
Livanova (b)(4) manufactures the centrifugal pump 5 (cp5).The event occurred in (b)(6).This medwatch report is being filed on behalf of livanova (b)(4).A livanova field service representative was dispatched to the facility to investigate.The service representative checked the functionality of the unit and the cables and was unable to reproduce the reported issue.The software was updated to rev.210 and the unit was returned to service.As the issue was not reproduced, a root cause could not be determined.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.Evaluated on-site by livanova technician.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SORIN CENTRIFUGAL PUMP 5 (CP5)
Type of Device
CARDIOPULMONARY BYPASS PUMP SPEED CONTROL
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich D 809 39
GM  D 80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key4747723
MDR Text Key5706478
Report Number9611109-2015-00126
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112225
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,health professional
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date 04/03/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-02-60
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/01/2015
Initial Date FDA Received04/29/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/03/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-