The thermacor 1200 infusion system was being used at (b)(6) hosp on (b)(6) 2015.The hosp stated that the cassette started leaking, and they changed over to a level one unit.However, the hosp bio-med investigated the issue and found the operator had not loaded the tubing correctly into the pump door area and the roller pump door had pinched the tubing, causing the leakage.No pt injury or extended surgery time was needed for changing out the unit.In-service training is being conducted on 5/1/2015 to ensure the operators understand proper loading of the cassette.
|