It was reported during a surgical procedure the surgeon could not successfully find a correct position for the fusion 150 device on the patient's atrium.He finally decided to stop the procedure.A few moments later, he noticed bleeding coming from circumflex artery.The surgeon noticed several dislodged electrodes protruding from the device's probe.The surgeon used sutures to repair the circumflex artery, causing the patient to remain on pump bypass an additional hour.
|
(b)(4).Upon receipt, the device was evaluated and was confirmed to have several dislodged electrodes.Electrode dislodgement is believed to be caused by using forceps or metal graspers, in the electrode area instead of the areas indicated, to reposition the device.Device met all electrical specifications.
|