• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - COSTA RICA (HEREDIA) INTELLATIP MIFI? XP TEMPERATURE ABLATION CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - COSTA RICA (HEREDIA) INTELLATIP MIFI? XP TEMPERATURE ABLATION CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER Back to Search Results
Model Number M004PM4500K20
Device Problem Bent (1059)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/05/2015
Event Type  malfunction  
Event Description
Same case as mfr id # 2134265-2015-02699.Reportable based on analysis completed on (b)(4) 2015.It was reported that tip damage occurred.The target lesion was located in the right atrium.During use of the intellatip mifi¿ xp temperature ablation catheter, the physician noted that the tip was bent between the second and third electrode.This occurred with a second intellatip mifi¿ xp temperature ablation catheter as well.The procedure was completed with a non-bsc catheter.However analysis revealed that the adhesive between the electrodes was broken.
 
Manufacturer Narrative
Device evaluated by manufacturer - the device has a kink at the distal section while in the neutral position at 14mm from the tip.Rings #1 and #2 have broken adhesive and fluids under them.The right curve is placed in the template shaded are, but the left curve is not placed in the template shaded area, the device failed the dimensional test.The handle was opened, and no issues were detected in that section.The device was dissected and the guide coil was inspected finding no issues on it.The distal section was dissected finding the center support kinked at 14mm.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLATIP MIFI? XP TEMPERATURE ABLATION CATHETER
Type of Device
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
Manufacturer (Section D)
BOSTON SCIENTIFIC - COSTA RICA (HEREDIA)
302 parkway
la aurora
heredia
CS 
Manufacturer (Section G)
BOSTON SCIENTIFIC - COSTA RICA (HEREDIA)
302 parkway
la aurora
heredia
CS  
Manufacturer Contact
linda leimer
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key4766434
MDR Text Key5778578
Report Number2134265-2015-02751
Device Sequence Number1
Product Code OAD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P020025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/16/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/13/2017
Device Model NumberM004PM4500K20
Device Catalogue NumberPM4500K2
Device Lot Number16623756
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/07/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/16/2015
Initial Date FDA Received05/12/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/23/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-