The gore engineers have evaluated the returned device.Their investigation showed following: the c15-285 specimen consisted of the delivery catheter without the endoprosthesis, deployment knob, or deployment line.The catheter was unremarkable.The c15-286 specimen consisted of the deployment knob with attached deployment line (approximately 180 cm length).The line had multiple knots, and the distal end was frayed (see c15-286 pic 1).The identified nature event code is believed to be partial deployment: deployment line stuck.Based on the device examination performed, no manufacturing anomalies were identified.
|