NIPRO DIAGNOSTICS, INC. TRUETRACK; BLOOD GLUCOSE SYSTEM Back to Search Results
Model Number TRUETRACK
Device Problem Low Test Results (2458)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/21/2015
Event Type  malfunction  
Event Description
Customer complaint of e-5 error.Customer feels well and does not required medical attention.Customer states e-5 error appear after sample is applied.Customer declined a blood test.Verified the strips expired 6/29/2017 unable to confirm the open date of test strips of the storage conditions.Customer stores strips in the kitchen.No adverse event reported.
 
Manufacturer Narrative
(b)(4).Returned meter evaluated: missing rubber cover over mode switch.Root cause: user mechanical stress.Returned test strips produced "low" readings.Most likely underlying root cause of malfunction noted in the complaint: black chemistry due to improper storage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUETRACK
Type of Device
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
NIPRO DIAGNOSTICS, INC.
2400 n.w. 55th ct.
fort lauderdale FL 33309
Manufacturer Contact
karen devincent
2400 n.w. 55th ct.
fort lauderdale, FL 33309
9546779201
MDR Report Key4771997
MDR Text Key5780449
Report Number1052693-2015-00712
Device Sequence Number1
Product Code CGA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K032657
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Remedial Action Replace
Type of Report Initial
Report Date 05/12/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/12/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberTRUETRACK
Device Catalogue NumberA4H01-81
Device Lot NumberRS4583
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer04/20/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received04/20/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-