BECKMAN COULTER AVANTI J-26 XP/XPI CENTRIFUGE; CENTRIFUGES (MICRO, ULTRA, REFRIGERATED) FOR CLINICAL USE Back to Search Results
Catalog Number 367502
Device Problems Device Emits Odor (1425); Sparking (2595)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/21/2015
Event Type  malfunction  
Event Description
The customer reported a burning smell and saw sparks from the avanti j26 xp/xpi centrifuge.There was no death or serious injury related to this event.
 
Manufacturer Narrative
A field service engineer (fse) evaluated the instrument on 4/21/2015.The fse found that the power line connector on the emi filter was burnt.The fse replaced the cable and the filter, which resolved the reported issue.(b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVANTI J-26 XP/XPI CENTRIFUGE
Type of Device
CENTRIFUGES (MICRO, ULTRA, REFRIGERATED) FOR CLINICAL USE
Manufacturer (Section D)
BECKMAN COULTER
250 south kraemer blvd.
brea CA 92821
Manufacturer (Section G)
BECKMAN COULTER
11800 sw 147th ave
miami FL 33196
Manufacturer Contact
miranda holland
11800 sw 147th ave
32-l23
miami, FL 33196
3053802031
MDR Report Key4774798
MDR Text Key5953030
Report Number3007448124-2015-00002
Device Sequence Number1
Product Code JQC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Health Professional
Type of Report Initial
Report Date 04/21/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/15/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number367502
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/21/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-