Sorin group received a report that during a vein harvesting procedure, the bipolar jaw broke off.The piece was retrieved from the patient's leg and discarded.The bipolar device was replaced and the procedure continued without any issues.There was no injury to the patient.One vascuclear precision bipolar device was returned to sorin group usa for further investigation.The device was subjected to visual inspection and functional testing.The visual inspection confirmed the user report that the upper jaw had broken off.The lower jaw was found to be undamaged.During testing, all functionalities of the device worked as expected.Based on the results of the evaluation of the returned device, it has been concluded that the issue was likely the result of advancing the bipolar with the jaws in the open position, contrary to the ifu.The vascuclear precision bipolar instructions for use states: "caution: do not advance or torque the instrument with the jaws open or use the jaws for spread dissection.These actions will damage the instrument.".
|