Model Number 8637-40
Device Problems Difficult to Interrogate (1331); Unstable (1667)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  Injury  
Event Description
It was reported that a healthcare professional (hcp) reported a suspected flipped pump which was later confirmed by x-ray.During a routine refill telemetry was difficult but they were eventually able to read the pump.The hcp reported that when palpated it felt like the pump port where the catheter connected was coming off the left of the pump.The hcp wondered if when the patient was being transferred using a hoyer lift it could be pulling or affecting the pump.The pump pocket was very loose.Surgical intervention was required to flip the pump back over and secure with sutures.The pump remained implanted and in service, and the patient was alive with no injury.The pump was used to infuse lioresal (baclofen).Follow up was being conducted to obtain additional information.If additional information is received a follow up report will be sent.
 
Manufacturer Narrative
Concomitant medical products: product id 8780, serial# (b)(4), implanted: (b)(6) 2014, product type: catheter.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHROMED II
Type of Device
PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MDT PUERTO RICO OPERATIONS CO
rd 31 km 24 hm 4
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC NEUROMODULATION
7000 central avenue ne rcw215
minneapolis MN 55432
Manufacturer Contact
diane wolf
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263987
MDR Report Key4782166
MDR Text Key5819798
Report Number3004209178-2015-09387
Device Sequence Number1
Product Code LKK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 04/28/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/14/2015
Device Model Number8637-40
Device Catalogue Number8637-40
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/28/2015
Initial Date FDA Received05/19/2015
Date Device Manufactured07/19/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age00039 YR
-
-