SYNTHES OBERDORF HAND SWITCH FOR ELECTRIC PEN DRIVE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 05.001.012 |
Device Problems
Sticking (1597); Difficult or Delayed Activation (2577)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 06/24/2014 |
Event Type
malfunction
|
Event Description
|
It was reported that prior to surgery during pre-testing, it was discovered that the hand switch for electric pen drive device had a sticky trigger.There were no delays to the planned surgical procedure as an identical spare device was available for use.There was no patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Manufacturer Narrative
|
The actual device was returned for evaluation.Reliability engineering evaluated the device and observed that the finger extension was damaged.Therefore, the reported condition was duplicated and confirmed.The assignable root cause was determined to be due to usage wear over time.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|