MAUDE Adverse Event Report: STRYKER SPINE-SWITZERLAND (MDR) XIA 3 TITANIUM POLYAXIAL SCREW DIA 6.5 X 45 MM; SPINAL INTERLAMINAL FIXATION ORTHOSIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER SPINE-SWITZERLAND (MDR) XIA 3 TITANIUM POLYAXIAL SCREW DIA 6.5 X 45 MM; SPINAL INTERLAMINAL FIXATION ORTHOSIS Back to Search Results
Catalog Number 482316545
Device Problems Fracture (1260); Difficult to Remove (1528); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/30/2015
Event Type  malfunction  
Event Description
It was reported that the patient came in for a revision and when removing the screw only half of it came out ended up skipping t4 level and extended it through the pelvis.
 
Manufacturer Narrative
Method: visual inspection; functional inspection; device history review; complaint history review; risk assessment; results: inspection of the device revealed beach marks on the fracture surface, which is consistent with fatigue fracture.Per the ifu, these devices are only meant to provide support during fusion and cannot be expected to indefinitely withstand the loads of normal bone.It was stated that fusion was not achieved.Manufacturing records were reviewed for the corresponding lot and no relevant issues were identified conclusion: the most likely cause of the customer reported event is fatigue fracture due to normal loads imposed by the body.
 
Event Description
It was reported that the patient came in for a revision and when removing the screw only half of it came out ended up skipping t4 level and extended it through the pelvis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIA 3 TITANIUM POLYAXIAL SCREW DIA 6.5 X 45 MM
Type of Device
SPINAL INTERLAMINAL FIXATION ORTHOSIS
Manufacturer (Section D)
STRYKER SPINE-SWITZERLAND (MDR)
le crêt-du-locle 10 a
-
la chaux-de-fonds 02300
CH  02300
Manufacturer (Section G)
STRYKER SPINE-SWITZERLAND (MDR)
le crêt-du-locle 10 a
-
la chaux-de-fonds 02300
CH   02300
Manufacturer Contact
linford leitch
2 pearl court
allendale, NJ 07401
2017608000
MDR Report Key4796422
MDR Text Key5819652
Report Number3005525032-2015-00063
Device Sequence Number1
Product Code KWP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K113666
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,health professional
Reporter Occupation Other
Type of Report Initial
Report Date 04/30/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number482316545
Device Lot NumberB34676
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/11/2015
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/30/2015
Initial Date FDA Received05/26/2015
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/25/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-