• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION EON MINI IPG, 16-CHANNEL RECHARGEABLE; SCS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION EON MINI IPG, 16-CHANNEL RECHARGEABLE; SCS IPG Back to Search Results
Model Number 3788
Device Problems Failure to Charge (1085); No Device Output (1435); Use of Device Problem (1670)
Patient Problem Inadequate Pain Relief (2388)
Event Type  Injury  
Event Description
It was reported, the patient was unable to establish communication between her ipg and external devices.The patient's programmer also reflected a communication error.The patient stated she has not charged the device in approximately 4 months.The sjm representative met with the patient and confirmed the issue.Subsequently, the patient underwent surgical intervention on (b)(6) 2015, where the ipg was explanted and replaced.The patient reported effective therapy postoperative.Please note the implant date and concomitant medical products are unknown.
 
Manufacturer Narrative
Recall: 1627487-12192011-003-r.This ipg serial number was included in field advisories.(b)(4).Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Manufacturer Narrative
Evaluation codes: results: the claim about inoperable ipg was confirmed.As received the ipg was non-responsive.The reported complaint cannot be analyzed as this is considered to be a user error.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EON MINI IPG, 16-CHANNEL RECHARGEABLE
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
plano TX
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key4806531
MDR Text Key5830543
Report Number1627487-2015-23306
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup
Report Date 06/23/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/28/2012
Device Model Number3788
Device Lot Number3190241
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/11/2015
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/23/2015
Initial Date FDA Received05/29/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received07/17/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/24/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487-07262012-002-R
Patient Sequence Number1
Patient Outcome(s) Other;
-
-