• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION LAMITRODE TRIPOLE 16C LEAD, 60 CM; SCS LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION LAMITRODE TRIPOLE 16C LEAD, 60 CM; SCS LEAD Back to Search Results
Model Number 3214
Device Problems Fracture (1260); Device Stops Intermittently (1599); Impedance Problem (2950)
Patient Problems Pain (1994); Inadequate Pain Relief (2388)
Event Date 05/07/2015
Event Type  Injury  
Event Description
The patient reported he experiences a painful jolt after which his scs system stimulation stops.An sjm representative met with the patient and ran a system diagnostic which revealed invalid impedances were present.X-rays were taken and indicated the patient's lead has fractured.Surgical intervention may take place at a later date to address the issue.
 
Manufacturer Narrative
(b)(4).Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAMITRODE TRIPOLE 16C LEAD, 60 CM
Type of Device
SCS LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key4817978
MDR Text Key19313072
Report Number1627487-2015-06250
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 05/15/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/04/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/16/2009
Device Model Number3214
Device Lot Number112931
Was Device Available for Evaluation? No
Date Manufacturer Received05/15/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/17/2007
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MODEL 3716, SCS IPG
Patient Outcome(s) Other;
Patient Age48 YR
-
-